Despatch MIC1462 Controller Manual Manual de usuario Pagina 81

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 112
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 80
77
EXIT FROM CALIBRATION MODE
To exit from Calibration Mode, cycle power to the instrument.
NOTE: An automatic exit is made from Calibration Mode if there is no key activity for
two minutes.
CALIBRATION CHECK
1. Set the Controller to the required configuration (using link jumpers and front panel
entry) as described in the appropriate Site Manual.
2. Power-up the Controller and leave it powered-up for at least five minutes (for RTD
and DC linear inputs) or at least 30 minutes (for thermocouple inputs).
3. After the appropriate delay for stabilization has elapsed, check the calibration by
connecting the appropriate input source and checking a number of cardinal points.
Vista de pagina 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 111 112

Comentarios a estos manuales

Sin comentarios