LAC SERIES OVEN
WITH PROTOCOL
PLUS
™
INSTRUCTION MANUAL
C-184
PN 143365
REVISION H
11/2007
Vista de pagina 0
1 2 3 4 5 6 ... 80 81

Indice de contenidos

Pagina 1 - INSTRUCTION MANUAL

LAC SERIES OVEN WITH PROTOCOL PLUS ™ INSTRUCTION MANUAL C-184 PN 143365 REVISION H 11/2007

Pagina 2 - Caution

1 INTRODUCTION This section provides an overview of the Despatch LAC Series forced air oven. The LAC Series Ovens have the most effective heat dist

Pagina 3

2 Specifications Dimensions LAC Model No. Chamber Size in (cm) Capacity feet3 (liters) Overall Size in (cm) Max. Number of Shelf Positions Exhaust

Pagina 4

3 Capacities LAC Model Number 1-10 1-38 A & B 1-67 2-12 2-18 Maximum Load Lbs Maximum Shelf Load Lbs Exhaust CFM Recirculating Fan CFM

Pagina 5

4 Power Line voltages may vary in some geographical locations. If your line voltage is much lower than the oven voltage rating, warm up time will b

Pagina 6 - PREFACE

5 THEORY OF OPERATION This section details the function and operation of assemblies and subassemblies on the Despatch LAC Series Ovens. These ovens

Pagina 7

6 Damper Control The oven is equipped with a manually adjustable damper mechanism. The damper control arm is located on the front panel of the oven

Pagina 8 - TABLE OF CONTENTS

7 Full Open Position When the damper is in the full open position, the chamber will operate at its minimum operating temperature. Friction heat fr

Pagina 9

8 INSTRUCTIONS The INSTRUCTIONS section provides directions for unpacking, installation, operation and maintenance of the LAC Series oven. Unpacki

Pagina 10 - Special Features

9 5. If any of these items are missing from the packaged contents, contact Despatch Industries to have the appropriate materials forwarded to you.

Pagina 11 - Specifications

10 Operating Users and operators of this oven must comply with operating procedures and training of operating personnel as required by the Occupatio

Pagina 12 - Temperature

Dear Customer, Thank you for choosing Despatch Industries. We appreciate the opportunity to work with you and to meet your heat processing

Pagina 13

11 The two shelves are designed to be pulled out about half way without tipping for loading and unloading. The support capacity of the shelves is li

Pagina 14 - THEORY OF OPERATION

12 Startup For fastest oven heat-up time, close the fresh-air vent. After the desired temperature is reached, the vent may be adjusted as needed.

Pagina 15 - Damper Control

13 CONTROL The special features of the Protocol PlusTM control include: PID tuning Ramp/Soak programming of up to 64 segments Segment loopin

Pagina 16 - Other Damper Settings

14 Protocol Plus Faceplate and Wiring Diagram

Pagina 17 - Unpacking and Inspection

15 Operating Modes The Protocol Plus control has five modes of operation available: Stopped Mode: All control and relay outputs are off. Stopped M

Pagina 18

16 High Limit The control has an integrated high limit function which will disable the heater output when tripped. If the high limit does trip, the

Pagina 19 - Operating

17 Displays The Protocol Plus control has two displays. A dedicated LED upper display shows the oven temperature. A two-line LCD lower display provi

Pagina 20 - Pre-Startup Checklist

18 Outputs The Protocol Plus control has seven different outputs available. Heating output: The control output is a DC voltage open-collector out

Pagina 21 - Startup

19 Relay (Continued) Use the Relay Card Optional Ay p/n 144562 to add relays to the standard controller. Each relay card contains two relays (maximu

Pagina 22 - Theory of Control Operation

20 Instructions Start-Up These instructions are provided as a quick reference for operating the Protocol Plus control. If the Profile Mode is to be

Pagina 24 - Fast Start Mode

21 Operation Manual Mode Press the Select key until Manual is displayed (note you can press the Run key at any time to activate Manual Mode). 1.

Pagina 25 - Indicators

22 Timer Mode 1. Press the Select key until Timer is displayed (note you can press the Run key at any time to activate Timer Mode). 2. Press the

Pagina 26 - Key Functions

23 Profile Mode 1. Press the Select key until Profile is displayed. “None” may be displayed if a profile has not been selected or no profiles ente

Pagina 27 - Outputs

24 Setup Mode Configuration of the control and programming of the ramp/soak profiles must be done in the Setup Mode. To access Setup Mode, the contr

Pagina 28 - Optional Software

25 Instructions for Setup Mode Pages Program Page Programming of the profiles is provided on the Program Page. Eight profiles are available with up

Pagina 29 - Instructions

26 Menu Item Display Description Ramp Time Seg 1 Pro-1 Seg-1 Ramp Time Ramp time for segment 1 of profile Event 1 Set Value* Pro-1 Seg-1 Ramp Event

Pagina 30 - Operation

27 Profile # There are eight profiles available. Segment# Recipe segments 1 through 8 may be programmed, each with its own set of events, ramp and

Pagina 31 - Timer Mode

28 Sample Profile Programming Table Profile Number____1______ Profile Name__________ Seg-ment Ramp Soak Time Events Temp-erature Time Events 1

Pagina 32 - Auto Start Mode

29 Auto Start Page (optional) If the optional real time clock has been installed, the Auto Start Page can be configured to automatically start Manua

Pagina 33

30 Menu Item Display Description Range Enable Autostart Auto Start Enable Enable (yes) or disable (no) the Autostart function No, Yes Sunday mode Au

Pagina 35

31 PID Page The PID Page contains parameters which control the response to the setpoint and process variable input. To access the PID Page, enter t

Pagina 36

32 Control Page The Control Page contains various parameters which control miscellaneous functions. To access the Control Page, enter the Setup Mode

Pagina 37 - Sample Profile

33 Communication Page (optional) The Communication Page contains parameters for the communications feature. To access the Communications Page, enter

Pagina 38 - Auto Start Page (optional)

34 Relay Outputs Page (optional) The Relay Outputs Page configures the four alarm/event outputs. Each output has a dedicated indicator light and can

Pagina 39

35 Test Page The Test Page contains parameters which allow manual control of the heat control and optional relay outputs and should be used only for

Pagina 40 - PID Page

36 Zone Calibration Page The Zone Calibration Page allows adjustment of the displayed temperature versus the actual temperature measured by the cont

Pagina 41 - Control Page

37 As an example, the control is displaying 400°F with the setpoint being 400°F, but the center of the oven chamber is actually 395°F. This can occu

Pagina 42

38 Sensor Calibration Page The Sensor Calibration Page has parameters which can change the internal calibration of the temperature sensor input sign

Pagina 43

39 19. Connect a calibration instrument with a type J thermocouple output to the control sensor input. 20. Connect a voltage measurement device to

Pagina 44 - Test Page

40 Enable Page The Enable Page controls access to the other Setup Pages. The setpoint minimum and maximum values, and the security passwords are als

Pagina 46

41 Digital Inputs (optional) The Protocol Plus control can be run by external inputs wired to the control from an external source such as a PLC or c

Pagina 47 - Sensor Calibration Page

42 Error Messages and Alarms The Alarm Status Hi-limit LED is flashing. This indicates a problem with the thermocouple, or the Hi-limit setpoint ha

Pagina 48

43 Quick Reference and Default Values Program Page Menu Item Display Default Range Setting Ramp Time Seg 1 Pro-1 Seg-1 Ramp Time 00:00 00m00s to 99h

Pagina 49 - Enable Page

44 Programming Table Profile Number__________ Profile Name__________ Seg-ment Ramp Soak Time Events Temp-erature Time Events 1 2 3 4 1 2 3 4 1

Pagina 50 - Digital Inputs (optional)

45 Autostart Menu Item Display Default Range Setting Enable Autostart Auto Start Enable No No, Yes Sunday mode Auto Start Sun Mode Off Off, Manual,

Pagina 51 - Error Messages and Alarms

46 Control Menu Item Display Default Range Setting Cycle Time Control Cycle Time 1 1 to 60 seconds High limit setpoint Control Hi-Lim SP Max HiLimSP

Pagina 52

47 Real Time Clock Menu Item Display Default Range Setting Day of the week Clock Day Mon Sun, Mon, Tue, Wed, Thu, Fri, Sat Time of day Clock HH:MM 0

Pagina 53

48 Test Menu Item Display Default Range Setting Heater output Test HeatOut Off On High limit relay Test HiLimOut Off On Relay 1 output Test Rly1 Ou

Pagina 54

49 Enable Page Menu Item Display Default Range Setting Profiles Enable Profile 1-8 Yes Yes or No Autostart Enable Auto Start No Yes or No PID Enabl

Pagina 55

50 Technical Specifications UL, cUL listed: UL file E136675 CE compliance to: EMC Directive 89-366/EEC European Standard EN55011/1991 European

Pagina 56

ii PREFACE This manual is your guide to the Despatch oven. It is organized to give you the information you need quickly and easily. The INTRODUCTI

Pagina 57

51 MAINTENANCE Do not attempt any service on this oven before opening the main power disconnect switch. Checklist  Keep equipment clean. Gra

Pagina 58

52 Tests Tests should be performed carefully and regularly. The safety of personnel as well as the condition of equipment may depend upon the prope

Pagina 59 - Technical Specifications

53 Replacement Parts To order or return parts, contact the Service Products Division at Despatch. The Service Products features our Response Cent

Pagina 60 - Checklist

54 Protocol Plus™ Instrument (Tools needed: 1/4" socket driver with #1 bit, #2 Phillips screwdriver) 1. Disconnect the power. 2. Remove the

Pagina 61

55 Fan Motor (Tools needed: Square recess driver w/#1 bit, 5/32 inch Allen wrench, one quarter (¼) inch socket set) 1. Remove the screws from the

Pagina 62 - Replacement

56 16. Replace and fasten the fan inlet cover. 17. Adjust the fan wheel for 3/16 inch clearance between the wheel and the inlet ring and tighten t

Pagina 63 - Heater Unit

57 TROUBLESHOOTING Equipment which operates for long periods of time may develop problems. Below are possible problems and suggested solutions. If

Pagina 64 - Fan Motor

58 Difficulty Probable Cause Suggested Remedy Excess surface or door temp. Door seal deterioration Replace door seal. Improper airflow Fan motor fai

Pagina 65

59 DRAWINGS LAC1-10-5 Excerpt from drawing 141852H1, 8-2000

Pagina 67

iii Revision B: Corrected Sensor Calibration Page instruction Revision C: Added definitions and sample profile to Program Page description Revis

Pagina 68 - LAC1-10-5

61 LAC1-38A-5 Excerpt from drawing 141853F1,4-2000

Pagina 69 - LAC1-10-5 Data

62 LAC1-38A-5 Data

Pagina 70 - LAC1-38A-5

63 LAC1-38B-5 Excerpt from drawing 141854G1, 4-2000

Pagina 72 - LAC1-38B-5

65 LAC1-67-5 Excerpt from drawing 141855G1, 4-2000

Pagina 74 - LAC1-67-5

67 LAC2-12-5 Excerpt from drawing 141856G1, 4-2000

Pagina 76 - LAC2-12-5

69 LAC2-18-5 (Excerpt from drawing 141857G1, 4-2000)

Pagina 78 - LAC2-18-5

iv TABLE OF CONTENTS INTRODUCTION ...

Pagina 79 - LAC2-18-5 Data

71 APPENDIX: Temperature Scale Conversion and Optional MRC5000 Setup Temperature Scale Conversion (C/F) The Protocol Plus controller can be operate

Pagina 80 - APPENDIX: Temperature Scale

72 Optional MRC5000 Recorder Setup The temperature is retransmitted to the Recorder from the Controller. Set up the Recorder as follows: 1. Make

Pagina 81